1.      ON/OFF key: Turns the instrument on and off. 

2.       TIME key: Allows the display of date and time, in the first line for about 8 seconds.

3.       SHIFT FNC key: Activates the Shortcut window. 

4.       Graphic display. 

5.       Function keys F1, F2, F3: Activate the function in the bottom line of the display. 

6.       ENTER key: In the menu, confirms the data entered. 

7.      ESC key: Allows exiting from the menu or, in case of a submenu, exiting from the current level display. 

8.       Navigation keys ▲▼: Allows navigation through the menus. 

9.       PRINT key: Starts and ends the data transfer to the serial/USB communication port. 

10.   MEM key: Starts and ends the recording of the data. 

11.   SETUP key: Allows entering and exiting the instrument’s functioning parameter setting menu. 

12.   SICRAM inputs for the probes. 

13.   Battery cover. 

14.   RS232 serial port. 

15.   USB port. 

16.   Power supply input.

TECHNICAL CHARACTERISTICS


 

Dimensions (Length x Width x Height)                    220x180x50 mm

Weight                                                                    1,100 g (batteries included)

Materials                                                                 ABS, polycarbonate and aluminum

Display                                                                    Backlit, Dot Matrix

                                                                               128x64 points, visible area 56x38mm

Operating conditions

Operating temperature                                             -5 … 50°C

Warehouse temperature                                           -25 … 65°C

Working relative humidity                                         0 … 90% RH without condensation

Protection degree                                                   IP64

 Instrument uncertainty                                            ± 1 digit @ 20°C

 Barometric pressure measurement with internal sensor

Measuring range                                                        600…1100hPa

Resolution                                                                  0.1hPa

Accuracy                                                                   0.5hPa

Response time                                                            1s

 Instrument temperature measurement with Pt100 probe

Pt100 measuring range                                              -200…+650°C

Resolution                                                                  0.01°C in the range ±199.99°C, 0.1°C in the remaining range

Accuracy                                                                   0.01°C in the range ±199.99°C,

                                                                                  0.1°C in the remaining range

Temperature drift @20°C                                          0.003%/°C

Drift after 1 year                                                        0.1°C/year

 Instrument relative humidity measurement (capacitive sensor)

Measuring range                                                        0…100%RH

Resolution                                                                  0.1%RH

 

Text Box:  

Accuracy                                                                   0.1%RH 

Temperature drift @20°C                                          0.02%RH/°C

Drift after 1 year                                                        0.1%RH/year

 Power

Mains adapter                                                           12Vdc/1A

Batteries                                                                    4 1.5V type C-BABY batteries

Autonomy                                                                  RH and temperature probes:

                                                                                 200 hours with 7800mAh alkaline batteries Hot-wire probe @ 5m/s:

                                                                                 100 hours with 7800mAh alkaline batteries

Power absorbed with instrument off                         < 45  A

Security of stored data                                             Unlimited

Connections

Input for probes with SICRAM module                    8x 8-pole male DIN45326 connector

 RS232C serial interface

Type                                                                          RS232C electrically isolated

Baud rate                                                                   Can be set from 1200 to 38400 baud

Data bit                                                                      8

Parity                                                                         None

Stop bit                                                                      1

Flow Control                                                             Xon/Xoff

Serial cable length                                                      Max. 15 m

 USB interface

Type                                                                           1.1 – 2.0 electrically isolated

 Memory                                                                     divided into 64 blocks.

 

Memory capacity                                                     67600 recordings for 8 inputs each.

 

Storage interval                                                       selectable among: 15, 30 seconds, 1, 2, 5,

                                                                                 10, 15, 20, 30 minutes and 1 hour.

 

Print interval                                                            selectable among: 15, 30 seconds, 1, 2, 5,

                                                                                  10, 15, 20, 30 minutes and 1 hour.

 EMC Standard Regulations

Safety                                                                          EN61000-4-2, EN61010-1 level 3

Electrostatic discharges                                                EN61000-4-2 level 3

Fast electric transients                                        EN61000-4-4 level 3, EN61000-4-5 level 3

Voltage variations                                               EN61000-4-11 Electromagnetic interference susceptibility                                    IEC1000-4-3 Electromagnetic

                                                                          interference emission  EN55020 class B

 

 